Local Coverage Determination (LCD)

Implantable Continuous Glucose Monitors (I-CGM)

L38662

Expand All | Collapse All
Proposed LCD
Proposed LCDs are works in progress that are available on the Medicare Coverage Database site for public review. Proposed LCDs are not necessarily a reflection of the current policies or practices of the contractor.

Document Note

Note History

Contractor Information

LCD Information

Document Information

Source LCD ID
N/A
LCD ID
L38662
Original ICD-9 LCD ID
Not Applicable
LCD Title
Implantable Continuous Glucose Monitors (I-CGM)
Proposed LCD in Comment Period
N/A
Source Proposed LCD
DL38662
Original Effective Date
For services performed on or after 11/02/2020
Revision Effective Date
For services performed on or after 11/04/2024
Revision Ending Date
N/A
Retirement Date
N/A
Notice Period Start Date
09/19/2024
Notice Period End Date
11/03/2024

CPT codes, descriptions, and other data only are copyright 2023 American Medical Association. All Rights Reserved. Applicable FARS/HHSARS apply.

Fee schedules, relative value units, conversion factors and/or related components are not assigned by the AMA, are not part of CPT, and the AMA is not recommending their use. The AMA does not directly or indirectly practice medicine or dispense medical services. The AMA assumes no liability for data contained or not contained herein.

Current Dental Terminology © 2023 American Dental Association. All rights reserved.

Copyright © 2024, the American Hospital Association, Chicago, Illinois. Reproduced with permission. No portion of the AHA copyrighted materials contained within this publication may be copied without the express written consent of the AHA. AHA copyrighted materials including the UB‐04 codes and descriptions may not be removed, copied, or utilized within any software, product, service, solution, or derivative work without the written consent of the AHA. If an entity wishes to utilize any AHA materials, please contact the AHA at 312‐893‐6816.

Making copies or utilizing the content of the UB‐04 Manual, including the codes and/or descriptions, for internal purposes, resale and/or to be used in any product or publication; creating any modified or derivative work of the UB‐04 Manual and/or codes and descriptions; and/or making any commercial use of UB‐04 Manual or any portion thereof, including the codes and/or descriptions, is only authorized with an express license from the American Hospital Association. The American Hospital Association (the "AHA") has not reviewed, and is not responsible for, the completeness or accuracy of any information contained in this material, nor was the AHA or any of its affiliates, involved in the preparation of this material, or the analysis of information provided in the material. The views and/or positions presented in the material do not necessarily represent the views of the AHA. CMS and its products and services are not endorsed by the AHA or any of its affiliates.

Issue

Issue Description

The proposed LCD modifies the coverage criteria for Implantable Continuous Glucose Monitors (I-CGMs) based on the best available evidence.  

Issue - Explanation of Change Between Proposed LCD and Final LCD

CMS National Coverage Policy

This LCD supplements but does not replace, modify or supersede existing Medicare applicable National Coverage Determinations (NCDs) or payment policy rules and regulations for Implantable Continuous Glucose Monitors (I-CGM). Federal statute and subsequent Medicare regulations regarding provision and payment for medical services are lengthy. They are not repeated in this LCD. Neither Medicare payment policy rules nor this LCD replace, modify or supersede applicable state statutes regarding medical practice or other health practice professions acts, definitions and/or scopes of practice. All providers who report services for Medicare payment must fully understand and follow all existing laws, regulations and rules for Medicare payment for Implantable Continuous Glucose Monitors (I-CGM) and must properly submit only valid claims for them. Please review and understand them and apply the medical necessity provisions in the policy within the context of the manual rules. Relevant CMS manual instructions and policies may be found in the following Internet-Only Manuals (IOMs) published on the CMS Web site:

IOM Citations:

  • CMS IOM 100-08, Medicare Program Integrity Manual,
    • Chapter 13, Section 13.5.4 Reasonable and Necessary Provision in an LCD

Social Security Act (Title XVIII) Standard References:

  • Title XVIII of the Social Security Act, Section 1862(a)(1)(A) states that no Medicare payment shall be made for items or services which are not reasonable and necessary for the diagnosis or treatment of illness or injury.
  • Title XVIII of the Social Security Act, Section 1862(a)(7). This section excludes routine physical examinations.

Coverage Guidance

Coverage Indications, Limitations, and/or Medical Necessity

Compliance with the provisions in this policy may be monitored and addressed through post payment data analysis and subsequent medical review audits.

History/Background and/or General Information

Covered Indications

I-CGMs are class III medical devices that require premarket approval by the FDA. In order to be considered reasonable and necessary, the FDA approved indication must include use as a therapeutic CGM.1 The FDA recently approved expanding the indications of an implantable CGM product to replace fingerstick blood glucose measurements for diabetes treatment decisions.

Therapeutic I-CGMs are considered reasonable and necessary by Medicare when all of the following coverage criteria (1-4) are met:

  1. The beneficiary has diabetes mellitus (Refer to the ICD-10 code list in the LCD-related Policy Article for applicable diagnoses); and,
  2. The beneficiary’s treating practitioner has concluded that the beneficiary (or beneficiary’s caregiver) has sufficient training using the I-CGM prescribed as evidenced by providing a prescription; and,
  3. The I-CGM is prescribed in accordance with its FDA indications for use; and,
  4. The beneficiary for whom a I- CGM is being prescribed, to improve glycemic control, meets at least one of the criteria below:
    1. The beneficiary is insulin-treated; or,
    2. The beneficiary has a history of problematic hypoglycemia with documentation of at least one of the following:
      1. Recurrent (more than one) level 2 hypoglycemic events (glucose <54mg/dL (3.0mmol/L)) that persist despite multiple (more than one) attempts to adjust medication(s) and/or modify the diabetes treatment plan; or,
      2. A history of one level 3 hypoglycemic event (glucose <54mg/dL (3.0mmol/L)) characterized by altered mental and/or physical state requiring third-party assistance for treatment of hypoglycemia

CGM Continued Coverage

Every six (6) months following the initial prescription of the CGM, the treating practitioner conducts an in-person or Medicare-approved telehealth visit with the beneficiary to document adherence to their CGM regimen and diabetes treatment plan.

Limitations

I-CGM devices will not be considered reasonable and necessary for short-term (72 hours to 1 week) diagnostic use.

Exception

For those beneficiaries who have previously met the coverage criteria for a non-implantable therapeutic continuous glucose monitor through the Medicare DME benefit and subsequently choose to switch to the implantable device, they may do so with a provider order. However, all other coverage criteria above must be fulfilled in order for Medicare payment.

Summary of Evidence

Diabetes mellitus is a chronic metabolic disease involving an underproduction or resistance to insulin, resulting in elevated blood glucose levels. The two most common types of diabetes are type 1 (T1DM) and type 2 (T2DM).2 T1DM comprises approximately 5-10% of diabetes cases, and describes an etiology in which an autoimmune response damages the insulin-producing beta cells of the pancreas, resulting in insufficient insulin production. Conversely, T2DM is characterized by insulin resistance.2,3 In T2DM, an individual’s insulin production fails to offset the deficit that is created by the resistance.2 T2DM constitutes an estimated 90%-95% of diabetes cases.3

Approximately 11.3% of Americans have diagnosed diabetes, and an additional 3.4% of Americans are estimated to have undiagnosed diabetes. The prevalence of diabetes is known to increase with age. In Medicare aged populations in the United States (≥ 65 years of age), 22.4% of the population has diagnosed diabetes, with an additional 4.7% estimated to have undiagnosed diabetes.4 Economic evaluations have estimated that diabetes annually accounts for $237 billion in direct US medical expenditure, and an additional $90 billion in lost productivity.5

Acute complications of diabetes include hypoglycemia, hyperglycemia, diabetic coma, and nonketotic hyperosmolar coma. Chronic hyperglycemia, resulting from poorly controlled diabetes, may result in serious and life-threatening damage, including dysfunction and failure of the eyes, kidneys, nervous system and cardiovascular system. The complications of diabetes mellitus are far less common and less severe in people who have well-controlled blood sugar levels.2

To prevent and/or delay the development of both short and long term complications of diabetes, the American Diabetes Association recommends person-centered care aimed at timely treatment decisions following evidence-based guidelines. Person-centered care goals include minimizing the progression of hyperglycemia with more intensive approaches for individuals at higher risk which include those with higher glucose levels (e.g., fasting blood glucose 110-125 mg/dL, 2 hour post-challenge glucose 173-199 mg/dL, A1C > 6.0%).6

Diabetes treatment relies on diet, exercise, lifestyle modifications, and in many, medications in order to keep their blood sugar under a stable and controlled level. Close glucose monitoring either multiple times a day or continuously may be warranted in those diabetics that are having difficulty maintaining that level.2,7

Continuous glucose monitoring (CGM) devices measure glucose via interstitial fluid instead of blood. CGM systems traditionally rely on a multi-part system that often involves: 1.) a sensor that is inserted into subcutaneous tissue; 2.) A receiver/mobile application consisting of a display and interface; 3.) a transmitter component that is attached to the sensor and worn externally.7,8 The sensors for these subcutaneous CGM systems need to be removed and replaced every 6- to 14-days depending upon the system.7

Despite the number of benefits that subcutaneous CGM offer in the diabetes mellitus (DM) management paradigm, some CGM studies have reported relatively high attrition rates that patients attributed to discomfort, contact dermatitis, issues changing the sensors, and physical interference with activities of daily living.9 The rationale for implantable continuous glucose monitoring (I-CGM) seeks to circumvent these negative patient experiences. Unlike with subcutaneous CGM, I-CGM involves a sensor that is surgically implanted for longer spans of time between replacement, allowing for skin barrier closure, and less frequent manipulation.

The primary aim of this summary of evidence is to determine if I-CGM performs with equivalent clinical validity, efficacy, safety, and patient perceived benefit when compared with subcutaneous CGM devices. The secondary aim of this summary of evidence is to evaluate the expected magnitude of benefit across the aforementioned outcomes resulting from the inclusion of I-CGM into the diabetes management paradigm.

At the time of this review, the Eversense E3 is the only device that has received clearance for use by the United States Food and Drug Administration (FDA). However, it is this Medicare Administrative Contractor’s expectation that additional devices are in the development and/or FDA approval process and soon may be publicly available. Hence, this policy is written to reflect coverage criteria and accompanying evidentiary review and analysis on I-CGM devices in an agnostic manner without the endorsement of any specific product.

Food and Drug Administration (FDA) Approvals

Eversense Continuous Glucose Monitoring System: https://www.accessdata.fda.gov/cdrh_docs/pdf16/P160048A.pdf

Literature Analysis

Clinical Validity

The accuracy of CGM devices is most commonly assessed by evaluating mean absolute relative difference (MARD). The MARD value represents the average percent difference between glucose readings that are concurrently obtained at multiple timepoints via an index test (I-CGM in this instance) and via an established reference standard test. MARD is then reported as a percentage, with lower values signifying a higher degree of test accuracy.10 For reference, one narrative review reported that the majority of CGM systems achieve an average MARD of between 9% to 14%, and that a MARD of < 10% had been discussed as a threshold for a CGM system that is accurate enough to inform insulin dosage decisions.11 Thirteen of the studies (14 publications) identified by the literature search evaluated the accuracy of I-CGM with MARD.12-25

Four of these 13 studies performed comparative analyses analyzing the accuracy of I-CGM compared with subcutaneous CGM systems.13,14,18,20

Fokkert and colleagues conducted a multi-test diagnostic cohort study among 23 athletes with T1DM. All 23 participants utilized an Eversense device, Free Style Libre (FSL) Flash Glucose Monitoring (FGM) system (subcutaneous CGM), and Free Style Libre Precision NeoPro strips (diagnostic standard). During normal activity among this sample of patients, Eversense performed with a MARD of 13% ± 6%, and the FSL FGM system had a MARD of 12% ± 5%; differences in these MARD values were not statistically quantified.18

One study conducted a three-way MARD analysis, comparing the accuracy of Eversense, the FSL Pro system (subcutaneous CGM), and the Dexcom G5 system (DG5) in a head-to-head fashion, with the StatStrip Xpress meter serving as the diagnostic standard. In this trial, 23 adults with T1DM utilized all 4 glucose monitoring devices over a 6-week timespan. Over the course of this trial, the authors reported that the MARD of Eversense was 14.8% ± 14.8%; FSL Pro performed with a MARD of 16.3% ± 15.4%, and the MARD of Libre Pro was 18.0% ± 17.9%. Jafri et al., then performed a pairwise comparison between these three MARD values, and determined that in this study, DG5 was significantly more accurate (P=0.004) than FSL Pro, and Eversense was significantly more accurate than both DG5 (P=0.008) and FSL Pro (P<0.000).20

Boscari and colleagues performed two separate trials comparing the MARD of Eversense with that of DG5. In the first trial, 11 adults with T1DM and HbA1c > 10% at baseline were enrolled, and Yellow Spring Instruments (YSI) STAT plus 2300 was used as the reference standard. Over the 1-week follow-up, the MARD of I-CGM (11.4% [IQR: 5.04% to 18.54%]) was statistically significantly less accurate (signed-rank test P<0.05) compared with that of Dexcom G5 (DG5) (7.91% (IQR: 4.14% to 14.30%).13 The second trial was designed as a crossover RCT study, in which 16 adults with T1DM and HbA1c < 10% underwent 12 weeks of DG5 and 12 weeks of Eversense, with self-monitoring of blood glucose (SMBG) serving as the reference test. Among this sample of 16 participants, the MARD of Eversense (12.27% ± 11.55%) was significantly more accurate (P<0.001) when compared with the MARD of DG5 (13.14% ± 14.76%).14

Nine single-test diagnostic accuracy studies were also identified; these publications assessed the MARD of I-CGM compared against a diagnostic reference standard (YSI 2300 STAT Plus, SMBG, or venous blood laboratory measurements). Among these 9 studies, the average MARD of I-CGM ranged between 8.8% to 12.3%.15-17,19,21-25

Clinical Efficacy

Glycated hemoglobin (HbA1c) is a key measure used to evaluate diabetes risk and morbidity.2 Although the American Diabetes Association recognizes that individual HbA1c normal ranges may vary, they recognize 7% as the standard threshold which most adult diabetics try to stay below.26 None of the 16 studies included in the literature review performed a comparative analysis of HbA1c change between I-CGM and subcutaneous CGM. However, 6 noncomparative studies meeting inclusion did assess HbA1c among patients using Eversense, including 3 that performed a pretest-posttest analysis. Among those 3 before-and-after trials, all 3 determined that mean HbA1c levels statistically significantly decreased between baseline (mean: 7.4% - 7.6%) and final study follow-up (mean: 6.9% - 7.19%).16,21,27 No statistical analyses were performed in the other 3 studies evaluating HbA1c, but all 3 noted a trend of HbA1c reduction following management with I-CGM.14,19,28

Time in Range (TIR) is an often-utilized surrogate outcome, used to evaluate the efficacy of diabetes management. TIR measures the percentage of time that CGM readings fall within the target blood glucose range (commonly defined as between 70 – 180 mg/dL).10,29 One study was identified that directly compared TIR between patients using I-CGM systems and those utilizing subcutaneous CGM systems.14 Boscari et al., 2022 conducted a crossover RCT trial in which patients underwent 12 weeks of management with Eversense I-CGM, and 12 weeks of management with the Dexcom G5 (DG5) system. Results of this study showed that on average, 71.14% ± 12.29% of the I-CGM group readings were in the target range, which was statistically significantly favored (P<0.028) when compared with the 66.99% ± 11.8% mean TIR exhibited by the DG5 group. Four additional studies also evaluated TIR, but only among patients managing DM with I-CGM.23,24,27,28 One of these 4 studies performed a pretest-posttest statistical analysis among 100 adults with T1DM. Irace et al., found that diabetes management with Eversense was associated with a significant improvement of TIR compared with baseline.27 For 3 of the 4 noncomparative trials evaluating mean TIR during management with Eversense, the mean TIR was between 62.3% - 69% 23,24,27; the forth study had multiple different cohorts based upon patient characteristics and could not be pooled.28

Patient Reported Outcome Metrics (PROMs)

Boscari et al., 2022 was the only publication that comparatively analyzed patient reported outcome metrics (PROMs) between I-CGM and subcutaneous CGM.14 In this crossover trial, all 16 patients underwent 12 weeks of management with the Eversense system, and 12 weeks with the DG5. Patient quality of life (QOL) was measured with the validated 17-item Diabetes Distress Scale (DDS), which asks respondents to evaluate each item with a 6-point Likert Scale ranging from 1 (no problem) to 6 (serious problem). Eversense was statistically significantly favored compared with DG5 for the total DDS score (I-CGM, 2.1 ± 1.1; DG5, 2.6 ± 1.4; P=0.009) as well as for the emotional burden domain score (I-CGM, 2.1 ± 1.0; DG5, 2.7 ± 1.4; P=0.01), regimen-related distress domain score (I-CGM, 2.1 ± 1.1; DG5, 2.7 ± 1.3; P=0.007), and the interpersonal distress score (I-CGM, 2.0 ± 1.1; DG5, 2.6 ± 1.2; P=0.003). Only the physician-related distress domain score was not significantly different (P=0.12) between DG5 and Eversense groups. Patient satisfaction with CGM devices were measured with the Diabetes Treatment Satisfaction Questionnaire (DTSQ), which is an 8-item survey that patients are to score between 0 (not satisfied) and 6 (extremely satisfied). The average total DTSQ score provided by respondents using Eversense (31.2 ± 4.3) was not significantly different (P=0.96) when compared with the mean DTSQ respondent score for the Dexcom G5 system (31.3 ± 3.9).14

Two noncomparative studies (3 publications) evaluated patient perceived change in QOL following diabetes management with I-CGM.12,21,28 Among the 3 publications, QOL was assessed with the 36-Item Short Form Health Survey (SF-36), 28-Item DDS, and the Audit of Diabetes Dependent Quality of Life (ADDQoL) questionnaire. Only one of these two trials performed a before-and-after statistical analysis, and the authors concluded that no significant change in QOL was observed over time.28 Three noncomparative studies evaluated patient satisfaction with varying PROMs.12,16,28 One trial found that 84% of respondents would choose to be re-inserted with Eversense for implantation cycle if given the opportunity.12 Christiansen and colleagues examined patient satisfaction with the validated CGM Satisfaction Scale Questionnaire, which asks patients to rate items on a Likert Scale ranging from 1 (lowest satisfaction/largest hassle) through 5 (most satisfaction/least hassle).16 This study found that among the 35 respondents, the mean overall satisfaction score with I-CGM was 3.9 ± 0.6. Among the respondents, only 20% of I-CGM users indicated that they would be unwilling to use I-CGM again after the study concluded. A third study found that the median DTSQ score over 180 days changed between a median of 0-2 points, which was not statistically quantified.28

Safety

None of the 16 primary studies meeting inclusion for the evidentiary review performed a comparative safety analysis between I-CGM and subcutaneous CGM.

However, 10 noncomparative studies assessed the safety profile of I-CGM. Six of the 10 studies considered the incidence of serious adverse events (SAEs), with all 6 reporting that no SAEs occurred.15,17,19,21,27,30 Additionally, two studies stated that no adverse events occurred at all during study follow-up.13,14 For the remaining 8 studies assessing, the heterogeneity of how adverse events were grouped together and reported precluded the pooling of event rates across studies. Generally, events were infrequent, and mostly consisted of transient skin irritation at the patch site, site infection, pain, and difficulty removing the sensor during first attempt.15,16,19,21,23,27,28,30

Professional Society Recommendations

While the American Diabetes Association (ADA), American Association of Clinical Endocrinology (AACE), Diabetes Canada Clinical Practice Guidelines, and National Institute for Health and Care Excellence (NICE) all have position/guideline recommendations supporting CGM, none of these specifically discuss if and when I-CGM should be considered in the treatment paradigm.6,31-33 The following position statements directly invoke I-CGM, but only discuss optimal patient selection criteria:

American Diabetes Association (ADA)In the safety subheading of a 2022 clinical practice guideline published by the ADA Professional Practice Committee, the authors state that contact dermatitis has been associated with all CGM devices, often linked to isobornyl acrylate. In some patients who have sensitivities to the tape used with subcutaneous CGM patches, the authors comment that I-CGM may provide an alternative that could preclude dermatitis events.34

European Association for the Study of Diabetes (EASD) A consensus report published jointly by the EASD and the ADA in 2021 states that I-CGM could be a clinical alternative for patients who have sensitivities to the adhesive that is required with subcutaneous CGM.35

American Association of Clinical Endocrinology (AACE) In 2021, the AACE issued a Grade C recommendation that clinicians should ensure patients undergoing treatment with CGM do not ingest any substances that could interfere with reliable interstitial glucose readings.32 In the evidence section that informs this recommendation, Grunberger et al., cite other CGM systems that are susceptible to inaccuracy when in the presence of acetaminophen, ascorbic acid, xylose, bilirubin, or uric acid. Conversely, this section reports that Eversense sensors are only vulnerable to inaccuracy in the presence of mannitol or tetracycline, consistent with the contraindications listed by the FDA packaging.36

Other Expert Opinion

Deiss et al., 2019 present a collection of consensus best practice recommendations by a panel of international clinicians who were involved with early Eversense clinical trials.37 Among the best practices, the following considerations are mentioned in relation to optimal patient selection criteria:

  • The authors note that the adhesive for Eversense does not contain isobornyl acrylate. Additionally, they also discuss how the daily changing of adhesive allows for regular skin care at the site. They conclude that both of these characteristics make I-CGM a viable alternative for patients with a history of skin reactions to adhesives.
  • Deiss et al., suggest that I-CGM is an ideal alternative to subcutaneous CGM for people who have an ongoing environmental risk to continued permanent breach of the skin, such as dust exposure.
  • Deiss and colleagues mention the need for mobile device compatibility with the app, as well as ability to navigate the app. They then comment that certain groups of patients may be unable to access or use this technology and would not be suitable candidates for I-CGM.
Analysis of Evidence (Rationale for Determination)

This updated evidentiary review is being conducted to evaluate whether the published evidence base continues to support the equivalency of implantable continuous glucose monitors (I-CGMs) and traditional subcutaneous continuous glucose monitors (CGMs) in the management of adult diabetes mellitus (DM). The evidence suggests that for adult patients managing DM, I-CGM performs with little to no difference in diagnostic accuracy, patient-reported efficacy outcomes, objective efficacy measures, and device/procedure-related adverse event rate when compared to subcutaneous CGM. Furthermore, the evidence suggests that the optimal patient selection criteria for I-CGM and for subcutaneous CGM are largely overlapping. Although limited in quality, this evidence base is sufficient to determine noninferiority/equivalency between I-CGM and subcutaneous CGM.

Additionally, although no direct evidence was identified that assessed I-CGM utilization among non-insulin treated (NIT) patients with diabetes complicated by hypoglycemia, there is still adequate evidence to infer equivalence with subcutaneous CGM systems among this population based upon the totality of evidence. Firstly, among the approved indications for I-CGM, the U.S. Food and Drug Administration (FDA) has cleared Eversense to “provide glucose trend information”, and to “provide alerts for the detection and prediction of episodes of low blood glucose (hypoglycemia) and high blood glucose (hyperglycemia)”.38 Also, although outside of the scope of the main evidentiary review, three publications were identified via bibliographic cross-referencing that evaluated subcutaneous CGM benefit among this subgroup of patients. One of these three trials concluded that among NIT patients with DM complicated by hypoglycemia, subcutaneous CGM systems resulted in a significantly greater reduction in HbA1c than SMBG.33 The other two studies suggested CGM systems were able to detect hypoglycemic events among non-insulin treated patients that were overwhelmingly undetectable by symptoms or SMBG.39,40

Proposed Process Information

Synopsis of Changes
Changes Fields Changed
N/A
Associated Information
Sources of Information
Bibliography
Open Meetings
Meeting Date Meeting States Meeting Information
N/A
Contractor Advisory Committee (CAC) Meetings
Meeting Date Meeting States Meeting Information
N/A
MAC Meeting Information URLs
N/A
Proposed LCD Posting Date
Comment Period Start Date
Comment Period End Date
Reason for Proposed LCD
Requestor Information
This request was MAC initiated.
Requestor Name Requestor Letter
View Letter
N/A
Contact for Comments on Proposed LCD

Coding Information

Bill Type Codes

Code Description

Please accept the License to see the codes.

N/A

Revenue Codes

Code Description

Please accept the License to see the codes.

N/A

CPT/HCPCS Codes

Please accept the License to see the codes.

N/A

ICD-10-CM Codes that Support Medical Necessity

Group 1

Group 1 Paragraph:

N/A

Group 1 Codes:

N/A

N/A

ICD-10-CM Codes that DO NOT Support Medical Necessity

Group 1

Group 1 Paragraph:

N/A

Group 1 Codes:

N/A

N/A

Additional ICD-10 Information

General Information

Associated Information

Please refer to the related Billing and Coding Article: Implantable Continuous Glucose Monitors (I-CGM) (A58127) for documentation and utilization requirements as applicable.

 

Sources of Information
N/A
Bibliography
  1. Department of Health and Human Services. Centers for Medicare & Medicaid Services. CMS Rulings. CMS-1682-R. Accessed August 16, 2023, https://www.cms.gov/Regulations-and-Guidance/Guidance/Rulings/CMS-Rulings-Items/CMS1682R
  2. Sapra A, Bhandari P. Diabetes. StatPearls Publishing. Accessed August 9, 2023, https://www.ncbi.nlm.nih.gov/books/NBK551501/
  3. Centers for Disease Control and Prevention (CDC). What is Diabetes? Accessed August 10, 2023, https://www.cdc.gov/diabetes/basics/diabetes.html
  4. Centers for Disease Control and Prevention (CDC). Prevalence of Both Diagnosed and Undiagnosed Diabetes. Accessed August 10, 2023,
  5. American Diabetes Association (ADA). Economic Costs of Diabetes in the U.S. in 2017. Diabetes Care. May 2018;41(5):917-928. doi:10.2337/dci18-0007
  6. American Diabetes Association (ADA). Standards of Care in Diabetes-2023 Abridged for Primary Care Providers. Clin Diabetes. Winter 2022;41(1):4-31. doi:10.2337/cd23-as01
  7. Solano L. Seven facts you need to know about continuous glucose monitors. Jaapa. Sep 1 2022;35(9):41-45. doi:10.1097/01.JAA.0000854536.09072.5e
  8. Adolfsson P, Parkin CG, Thomas A, Krinelke LG. Selecting the Appropriate Continuous Glucose Monitoring System - a Practical Approach. Eur Endocrinol. Apr 2018;14(1):24-29. doi:10.17925/ee.2018.14.1.24
  9. Cowart K. A Review of the First Long-term Implantable Continuous Glucose Monitoring System Available in the United States. J Diabetes Sci Technol. Jan 2021;15(1):160-166. doi:10.1177/1932296819890865
  10. Heinemann L, Schoemaker M, Schmelzeisen-Redecker G, et al. Benefits and Limitations of MARD as a Performance Parameter for Continuous Glucose Monitoring in the Interstitial Space. J Diabetes Sci Technol. Jan 2020;14(1):135-150. doi:10.1177/1932296819855670
  11. Freckmann G. Basics and use of continuous glucose monitoring (CGM) in diabetes therapy. 2020;44(2):71-79. Journal of Laboratory Medicine. doi:doi:10.1515/labmed-2019-0189
  12. Barnard KD, Kropff J, Choudhary P, et al. Acceptability of Implantable Continuous Glucose Monitoring Sensor. J Diabetes Sci Technol. May 2018;12(3):634-638. doi:10.1177/1932296817735123
  13. Boscari F, Vettoretti M, Amato AML, et al. Comparing the accuracy of transcutaneous sensor and 90-day implantable glucose sensor. Nutr Metab Cardiovasc Dis. Feb 8 2021;31(2):650-657. doi:10.1016/j.numecd.2020.09.006
  14. Boscari F, Vettoretti M, Cavallin F, et al. Implantable and transcutaneous continuous glucose monitoring system: a randomized cross over trial comparing accuracy, efficacy and acceptance. J Endocrinol Invest. Jan 2022;45(1):115-124. doi:10.1007/s40618-021-01624-2
  15. Christiansen MP, Klaff LJ, Bailey TS, Brazg R, Carlson G, Tweden KS. A Prospective Multicenter Evaluation of the Accuracy and Safety of an Implanted Continuous Glucose Sensor: The PRECISION Study. Diabetes Technol Ther. May 2019;21(5):231-237. doi:10.1089/dia.2019.0020
  16. Christiansen MP, Klaff LJ, Brazg R, et al. A Prospective Multicenter Evaluation of the Accuracy of a Novel Implanted Continuous Glucose Sensor: PRECISE II. Diabetes Technol Ther. Mar 2018;20(3):197-206. doi:10.1089/dia.2017.0142
  17. Dehennis A, Mortellaro MA, Ioacara S. Multisite Study of an Implanted Continuous Glucose Sensor Over 90 Days in Patients With Diabetes Mellitus. J Diabetes Sci Technol. Jul 29 2015;9(5):951-6. doi:10.1177/1932296815596760
  18. Fokkert M, van Dijk PR, Edens MA, et al. Performance of the Eversense versus the Free Style Libre Flash glucose monitor during exercise and normal daily activities in subjects with type 1 diabetes mellitus. BMJ Open Diabetes Res Care. Aug 2020;8(1)doi:10.1136/bmjdrc-2020-001193
  19. Garg SK, Liljenquist D, Bode B, et al. Evaluation of Accuracy and Safety of the Next-Generation Up to 180-Day Long-Term Implantable Eversense Continuous Glucose Monitoring System: The PROMISE Study. Diabetes Technol Ther. Feb 2022;24(2):84-92. doi:10.1089/dia.2021.0182
  20. Jafri RZ, Balliro CA, El-Khatib F, et al. A Three-Way Accuracy Comparison of the Dexcom G5, Abbott Freestyle Libre Pro, and Senseonics Eversense Continuous Glucose Monitoring Devices in a Home-Use Study of Subjects with Type 1 Diabetes. Diabetes Technol Ther. Nov 2020;22(11):846-852. doi:10.1089/dia.2019.0449
  21. Kropff J, Choudhary P, Neupane S, et al. Accuracy and Longevity of an Implantable Continuous Glucose Sensor in the PRECISE Study: A 180-Day, Prospective, Multicenter, Pivotal Trial. Diabetes Care. Jan 2017;40(1):63-68. doi:10.2337/dc16-1525
  22. Mortellaro M, DeHennis A. Performance characterization of an abiotic and fluorescent-based continuous glucose monitoring system in patients with type 1 diabetes. Biosens Bioelectron. Nov 15 2014;61:227-31. doi:10.1016/j.bios.2014.05.022
  23. Sanchez P, Ghosh-Dastidar S, Tweden KS, Kaufman FR. Real-World Data from the First U.S. Commercial Users of an Implantable Continuous Glucose Sensor. Diabetes Technol Ther. Dec 2019;21(12):677-681. doi:10.1089/dia.2019.0234
  24. Tweden KS, Deiss D, Rastogi R, Addaguduru S, Kaufman FR. Longitudinal Analysis of Real-World Performance of an Implantable Continuous Glucose Sensor over Multiple Sensor Insertion and Removal Cycles. Diabetes Technol Ther. May 2020;22(5):422-427. doi:10.1089/dia.2019.0342
  25. Wang X, Ioacara S, DeHennis A. Long-Term Home Study on Nocturnal Hypoglycemic Alarms Using a New Fully Implantable Continuous Glucose Monitoring System in Type 1 Diabetes. Diabetes Technol Ther. Nov 2015;17(11):780-6. doi:10.1089/dia.2014.0375
  26. American Diabetes Association (ADA). Understanding A1C. A1c Does it All. Accessed August 16, 2023, https://diabetes.org/diabetes/a1c
  27. Irace C, Cutruzzolà A, Nuzzi A, et al. Clinical use of a 180-day implantable glucose sensor improves glycated haemoglobin and time in range in patients with type 1 diabetes. Diabetes Obes Metab. Jul 2020;22(7):1056-1061. doi:10.1111/dom.13993
  28. Renard E, Riveline JP, Hanaire H, Guerci B. Reduction of clinically important low glucose excursions with a long-term implantable continuous glucose monitoring system in adults with type 1 diabetes prone to hypoglycaemia: the France Adoption Randomized Clinical Trial. Diabetes Obes Metab. May 2022;24(5):859-867. doi:10.1111/dom.14644
  29. American Diabetes Association (ADA). CGM & Time in Range. Accessed June 6, 2023, https://diabetes.org/tools-support/devices-technology/cgm-time-in-range#:~:text=What%20is%20time%20in%20range,mg%2FdL%20for%20most%20people.
  30. Deiss D, Irace C, Carlson G, Tweden KS, Kaufman FR. Real-World Safety of an Implantable Continuous Glucose Sensor Over Multiple Cycles of Use: A Post-Market Registry Study. Diabetes Technol Ther. Jan 2020;22(1):48-52. doi:10.1089/dia.2019.0159
  31. Fonseca VA, Grunberger G, Anhalt H, et al. CONTINUOUS GLUCOSE MONITORING: A CONSENSUS CONFERENCE OF THE AMERICAN ASSOCIATION OF CLINICAL ENDOCRINOLOGISTS AND AMERICAN COLLEGE OF ENDOCRINOLOGY. Endocr Pract. Aug 2016;22(8):1008-21. doi:10.4158/ep161392.Cs
  32. National Institute for Health and Care Excellence (NICE). Type 2 diabetes in adults: management. Accessed August 16, 2023, https://www.nice.org.uk/guidance/ng28
  33. Wada E, Onoue T, Kobayashi T, et al. Flash glucose monitoring helps achieve better glycemic control than conventional self-monitoring of blood glucose in non-insulin-treated type 2 diabetes: a randomized controlled trial. BMJ Open Diabetes Res Care. Jun 2020;8(1)doi:10.1136/bmjdrc-2019-001115
  34. American Diabetes Association Professional Practice Committee. 7. Diabetes Technology: Standards of Medical Care in Diabetes-2022. Diabetes Care. Jan 1 2022;45(Suppl 1):S97-s112. doi:10.2337/dc22-S007
  35. Holt RIG, DeVries JH, Hess-Fischl A, et al. The Management of Type 1 Diabetes in Adults. A Consensus Report by the American Diabetes Association (ADA) and the European Association for the Study of Diabetes (EASD). Diabetes Care. 2021;44(11):2589-2625. doi:10.2337/dci21-0043
  36. Grunberger G, Sherr J, Allende M, et al. American Association of Clinical Endocrinology Clinical Practice Guideline: The Use of Advanced Technology in the Management of Persons With Diabetes Mellitus. Endocr Pract. Jun 2021;27(6):505-537. doi:10.1016/j.eprac.2021.04.008
  37. Deiss D, Szadkowska A, Gordon D, et al. Clinical Practice Recommendations on the Routine Use of Eversense, the First Long-Term Implantable Continuous Glucose Monitoring System. Diabetes Technol Ther. May 2019;21(5):254-264. doi:10.1089/dia.2018.0397
  38. U.S. Food & Drug Administration (FDA). Approval Order - Eversense Continuous Glucose Monitoring System. Accessed August 16, 2023, https://www.accessdata.fda.gov/cdrh_docs/pdf16/P160048A.pdf
  39. Gehlaut RR, Dogbey GY, Schwartz FL, Marling CR, Shubrook JH. Hypoglycemia in Type 2 Diabetes—More Common Than You Think: A Continuous Glucose Monitoring Study. J Diabetes Sci Technol. Apr 27 2015;9(5):999-1005. doi:10.1177/1932296815581052
  40. Munshi MN, Segal AR, Suhl E, et al. Frequent hypoglycemia among elderly patients with poor glycemic control. Arch Intern Med. Feb 28 2011;171(4):362-4. doi:10.1001/archinternmed.2010.539
  41.  

Revision History Information

Revision History Date Revision History Number Revision History Explanation Reasons for Change
11/04/2024 R8

Revision Effective: 10/03/2024

Revision Explanation: Annual review, no changes.

  • Other (Annual Review )
11/04/2024 R7

Revision Effective: 11/04/2024

Revision Explanation: Released policy to notice after open comment period with related articles.

  • Provider Education/Guidance
10/05/2023 R6

R5

Revision Effective: 10/05/2023

Revision Explanation: Annual review, no changes.

  • Other (Annual Review)
05/05/2022 R5

R4

Revision Effective: 01/05/2023

Revision Explanation: Annual review, added the extended approval details to the Coverage Indications, Limitations and/or Medical Necessity section.

  • Other (Annual Review)
05/05/2022 R4

Correcting nae for Rev 3

R3

Revision Effective: 05/05/2022

Revision Explantion: Corrected link in the indication and analysis of evidence for the FDA Eversense GCM information as the previous link had been moved.

  • Typographical Error
05/05/2022 R3

R3

Revision Effective: 05/05/2022

Revision Explantion: Corrected link in the indication and analysis of evidence for the FDA Efersense GCM information as the previous link had been moved.

  • Provider Education/Guidance
03/17/2022 R2

R2
Revision Effective: 03/17/2022
Revision Explanation: Updates to multiple references within the Bibliography section.

  • Provider Education/Guidance
09/30/2021 R1

R1
Revision Effective: 09/30/2021
Revision Explanation: Annual review, no changes were made.

  • Other (Annual Review)
N/A

Associated Documents

Attachments
N/A
Related National Coverage Documents
N/A
Public Versions
Updated On Effective Dates Status
09/27/2024 11/04/2024 - N/A Currently in Effect You are here
09/14/2024 11/04/2024 - N/A Superseded View
09/29/2023 10/05/2023 - 11/03/2024 Superseded View
Some older versions have been archived. Please visit the MCD Archive Site to retrieve them.

Keywords

N/A

Read the LCD Disclaimer