Local Coverage Determination (LCD)

Ambulatory Electrocardiograph (AECG) Monitoring

L39492

Expand All | Collapse All
Proposed LCD
Proposed LCDs are works in progress that are available on the Medicare Coverage Database site for public review. Proposed LCDs are not necessarily a reflection of the current policies or practices of the contractor.

Document Note

Note History

Contractor Information

LCD Information

Document Information

Source LCD ID
N/A
LCD ID
L39492
Original ICD-9 LCD ID
Not Applicable
LCD Title
Ambulatory Electrocardiograph (AECG) Monitoring
Proposed LCD in Comment Period
N/A
Source Proposed LCD
DL39492
Original Effective Date
For services performed on or after 06/11/2023
Revision Effective Date
N/A
Revision Ending Date
N/A
Retirement Date
N/A
Notice Period Start Date
04/27/2023
Notice Period End Date
06/10/2023

CPT codes, descriptions, and other data only are copyright 2023 American Medical Association. All Rights Reserved. Applicable FARS/HHSARS apply.

Fee schedules, relative value units, conversion factors and/or related components are not assigned by the AMA, are not part of CPT, and the AMA is not recommending their use. The AMA does not directly or indirectly practice medicine or dispense medical services. The AMA assumes no liability for data contained or not contained herein.

Current Dental Terminology © 2023 American Dental Association. All rights reserved.

Copyright © 2024, the American Hospital Association, Chicago, Illinois. Reproduced with permission. No portion of the AHA copyrighted materials contained within this publication may be copied without the express written consent of the AHA. AHA copyrighted materials including the UB‐04 codes and descriptions may not be removed, copied, or utilized within any software, product, service, solution, or derivative work without the written consent of the AHA. If an entity wishes to utilize any AHA materials, please contact the AHA at 312‐893‐6816.

Making copies or utilizing the content of the UB‐04 Manual, including the codes and/or descriptions, for internal purposes, resale and/or to be used in any product or publication; creating any modified or derivative work of the UB‐04 Manual and/or codes and descriptions; and/or making any commercial use of UB‐04 Manual or any portion thereof, including the codes and/or descriptions, is only authorized with an express license from the American Hospital Association. The American Hospital Association (the "AHA") has not reviewed, and is not responsible for, the completeness or accuracy of any information contained in this material, nor was the AHA or any of its affiliates, involved in the preparation of this material, or the analysis of information provided in the material. The views and/or positions presented in the material do not necessarily represent the views of the AHA. CMS and its products and services are not endorsed by the AHA or any of its affiliates.

Issue

Issue Description

Ongoing claims analysis indicates aberrant utilization of cardiac monitoring services. Therefore, this LCD defines the various types of AECG monitoring devices and the medical necessity of monitoring for improved patient care outcomes.

Issue - Explanation of Change Between Proposed LCD and Final LCD

In response to the comments that were received, the table listing the types of cardiac monitors has been deleted and placed in the related billing and coding article, several covered indications have been added to the Covered Indications section, and the use of hand/wrist held or smart phone based devices (that do not have FDA clearance) has been added to the Limitations section. Additionally, the definition of 24-hour monitoring station has been revised to provide clarity. Three sources have been added to the bibliography with a summary of one of the sources added to the Summary of Evidence.

CMS National Coverage Policy

This LCD supplements but does not replace, modify, or supersede existing Medicare applicable National Coverage Determinations (NCDs) or payment policy rules and regulations for ambulatory electrocardiograph monitors. Federal statute and subsequent Medicare regulations regarding provision and payment for medical services are lengthy. They are not repeated in this LCD. Neither Medicare payment policy rules nor this LCD replace, modify, or supersede applicable state statutes regarding medical practice or other health practice professions acts, definitions and/or scopes of practice. All providers who report services for Medicare payment must fully understand and follow all existing laws, regulations, and rules for Medicare payment for ambulatory electrocardiograph monitors and must properly submit only valid claims for them. Please review and understand them and apply the medical necessity provisions in the policy within the context of the manual rules. Relevant CMS manual instructions and policies may be found in the following Internet-Only Manuals (IOMs) published on the CMS Web site:

IOM Citations:

  • CMS IOM Publication 100-02, Medicare Benefit Policy Manual,
    • Chapter 15, Section 80 Requirements for Diagnostic X-Ray, Diagnostic Laboratory, and Other Diagnostic Tests
  • CMS IOM Publication 100-03, Medicare National Coverage Determinations (NCD) Manual,
    • Chapter 1, Part 1, Section 20.15 Electrocardiographic Services 
  • CMS IOM Publication 100-08, Medicare Program Integrity Manual,
    • Chapter 13, Section 13.5.4 Reasonable and Necessary Provision in an LCD


Social Security Act (Title XVIII) Standard References:

  • Title XVIII of the Social Security Act, Section 1861(s)(1), is related to physicians’ services
  • Title XVIII of the Social Security Act, Section 1861(s)(3), is related to outpatient diagnostic services
  • Title XVIII of the Social Security Act, Section 1862(a)(1)(A) states that no Medicare payment may be made for items or services which are not reasonable and necessary for the diagnosis or treatment of illness or injury
  • Title XVIII of the Social Security Act, Section 1862(a)(7). This section excludes routine physical examinations


Code of Federal Regulations (CFR) References:

  • CFR, Title 42, Volume 2, Chapter IV, Part 410.32(d)(3) Diagnostic x-ray tests, diagnostic laboratory tests, and other diagnostic tests: Conditions
  • CFR, Title 42, Volume 2, Chapter IV, Part 410.33 Independent diagnostic testing facility
  • CFR, Title 42, Volume 2, Chapter IV, Part 414.50 Physician or other supplier billing for diagnostic tests performed or interpreted by a physician who does not share a practice with the billing physician or other supplier

Coverage Guidance

Coverage Indications, Limitations, and/or Medical Necessity

Compliance with the provisions in this LCD may be monitored and addressed through post payment data analysis and subsequent medical review audits.

History/Background and/or General Information

Ambulatory Electrocardiograph (AECG) diagnostic procedures provide a record of the heart rhythm during daily activities. AECG can often identify the existence and determine the frequency of clinically significant rhythm disturbances and waveform abnormalities that are missed on a standard electrocardiogram (ECG).

AECG continues to advance at a rapid pace and incorporates various monitoring devices and indications for use. These devices range from Holter Monitors, Event Recorders/Monitors, Patch Recorders, External Loop Recorders, Mobile Cardiovascular Telemetry, Mobile Cardiovascular Outpatient Telemetry, Insertable Cardiac Monitor/Internal Loop Recorders, and Subcutaneous Cardiac Rhythm Monitors. They are differentiated by capabilities and the varying technical components among devices, such as intermittent/continuous recording, patient/rhythm activated recording, number of leads, time frames of use and whether the rhythm is interpreted in real time (attended surveillance) or if it is transferred from the device and interpreted at a later time. Regardless of the equipment, these devices provide valuable data for diagnosis of palpitations, syncope and other signs and symptoms suggestive of cardiac arrhythmia.

The scope of this LCD is to define the specific patient indications that support medical necessity for this type of monitoring.

Definitions

Acute Coronary Syndrome (ACS) – applies to patients with suspicion or confirmation of acute myocardial ischemia or infarction. It describes a large array of signs and symptoms that range from atypical chest discomfort, nonspecific electrocardiographic changes, normal cardiac biomarkers, non-ST-elevation myocardial infarction (NSTEMI), large ST-segment elevation myocardial infarction (STEMI), and cardiogenic shock.1

Silent Myocardial Ischemia – is defined as ‘objective documentation of myocardial ischemia in the absence of angina or anginal equivalents.’2

24-hour Monitoring Station (applicable only to mobile cardiac telemetry) - is defined as a facility in operation 24-hours per day, seven days a week, with a receiving station staffed with EKG technicians or other non-physician staff on a 24-hour per day basis. The technicians should have immediate 24-hour access to a physician to review transmitted data that falls outside of set parameters. The technicians should know how to contact available facilities to assist in any cardiac emergencies.3,4,5.

Covered Indications

AECG monitoring, when performed with a device that has FDA clearance, will be considered medically reasonable and necessary in the following situations:

  1. A standard 12-lead electrocardiograph (ECG), complete cardiac history and cardiac exam has not satisfactorily explained the patient's cardiac complaints and AECG testing will provide diagnostic information that will assist in developing a treatment plan or changing a treatment plan for patients that are at risk for cardiac arrythmias.3,6,7
    OR
  2. For patients experiencing unexplained syncope, near syncope, episodic dizziness, chest pain, palpitations, and/or shortness of breath.6
    OR
  3. To assess documented or suspected bradycardia.8
    OR
  4. For patients experiencing nocturnal arrhythmias.6
    OR
  5. To assess the average heart rate and adequacy of rate control in a patient with atrial fibrillation.6
    OR
  6. To aid in the regulation of anti-arrhythmic drug dosage.3,6,7,9
    OR
  7. To assess the effectiveness of arrhythmia therapy (e.g., post ablation).6
    OR
  8. To evaluate prognosis following acute coronary syndrome.6,10
    OR
  9. Pre/Post implantable cardiac defibrillator reprogramming.7
    OR
  10. To assess for silent myocardial ischemia in a patient with known or suspected coronary heart disease.3,6
    OR
  11. To assess for asymptomatic ventricular premature beats or non-sustained ventricular tachycardia in patients with hypertrophic cardiomyopathy, arrhythmogenic right ventricular cardiomyopathy, long QT syndrome, dilated or restrictive cardiomyopathy, congenital heart disease, or Brugada syndrome.6,7
    OR
  12. To evaluate for occult atrial fibrillation (A-Fib) as a potential cause of cardio-embolism in patients with cryptogenic stroke.4,6

Note: A 24-48 hour monitor is most appropriate for patients with daily or near daily symptoms.6 Otherwise, providers should order the most appropriate type of AECG for the patient based on their evaluation of the patient, the patient’s symptoms and the devices’ labeled indications.

Note: The duration of monitoring should be consistent with the patient’s signs and symptoms.3,6,11,12,13

Limitations

The following are considered not medically reasonable and necessary:

  1. Devices that do not have FDA clearance, including non-physician prescribed; hand/wrist held – smart phone based devices.
  2. AECG monitoring of potentially harmful ventricular arrhythmias requiring inpatient level of care.12,14
  3. Any 24-hour monitoring station that does not meet the definition in the definition section.3,4,5

Please refer to NCD 20.15 for additional indications and limitations.

Provider Qualifications

Services will be considered medically reasonable and necessary when all aspects of care are within the scope of practice of the provider’s professional licensure, when performed according to the supervision requirements per state scope of practice laws, and when all procedures are performed by appropriately trained providers in the appropriate setting.

Notice: Services performed for any given diagnosis must meet the indications and limitations stated in this LCD, the general requirements for medical necessity as stated in CMS payment policy manuals, any and all existing CMS national coverage determinations, and all Medicare payment rules.

Summary of Evidence

Clinical Practice

Zimetbaum et al14 provides an overview and evaluation for care with patients that present with palpitations. There are many factors to be considered in the etiology of palpitations. Cardiac disorders are the most common cause of palpitations and include arrhythmias, structural heart disease, conduction system abnormalities, valvular disease, pacemaker syndrome, atrial myoma, sympathetic stimulation and catecholamine release during exercise and high output cardiac states. Other etiology of palpitations may include psychiatric disorders, medication effects, substance abuse, endocrine disorders, and metabolic disorders.

The workup for a patient who presents with palpitations typically includes an extensive medical history along with a 12-lead ECG, physical exam and limited laboratory testing.14(p3)

When an arrhythmia is not identified on the 12-lead ECG, the patient’s medical history can be helpful to note the age of symptom onset, duration of palpitations, heart rate and rhythm regularity, additional symptoms that are associated with the palpitations (pre-syncope or syncope), the onset and resolution of the palpitations, effects of positional changes and if the patient is able to alleviate the palpitations on their own. In addition, the cardiac exam should include the patient’s family cardiac history and co-existing medical conditions such as pregnancy, metabolic and endocrine issues, chronic obstructive pulmonary disease, and psychiatric disorders.14(p2)

Following an extensive cardiac exam, if the etiology of the palpitations remains unclear limited laboratory testing for anemia, hyperthyroidism, and toxicology testing may be indicated, as well as echocardiography for patients with possible structural heart disease.

Holter vs. Continuous Long-Term Electrocardiographic (LT-ECG) Monitors

Barrett et al15 conducted a prospective analysis of cardiac arrhythmias using the Holter monitor and the Zio patch between April 2012 and July 2012. The goal was to compare arrhythmia detections over the course of 14 days to the arrhythmia events detected in the first 24 hours of monitoring.

One hundred and forty-six patients between the ages of 22 and 94 participated in the study. The average wear time of the AECG monitors ranged between one day for the Holter monitor and eleven days for the patch monitor. Both devices were activated simultaneously and monitored for 24 hours for the Holter monitor and up to 14 days for the patch monitor. All arrhythmias were evaluated by the Scripps Translational Science Institute and the McNemar’s test was implemented to compare if arrhythmias were clinically significant and what type of arrhythmias were detected by each monitor.

Arrhythmia detection was placed into six categories: supraventricular tachycardia (SVT) (>4 beats, not including A-Fib or A-Flutter), A-Fib/A-Flutter (>4 beats), pause >3 seconds, atrioventricular block (Mobitz type II or third-degree atrioventricular [AV] block), ventricular tachycardia (VT) (>4 beats), or polymorphic ventricular tachycardia/ventricular fibrillation.15(p14)

The results demonstrated that the “patch monitor detected 96 arrhythmia events compared with 61 arrhythmia events by the Holter monitor (P< .001)."15(p12) The patch monitor and Holter monitor detected the same number of events, but “14 clinically significant arrhythmia events were detected by the adhesive patch monitor that went undetected by the Holter monitor."15(p12)

The secondary effect was to determine the arrhythmias that were detected in the first 24 hours. It was noted that the Holter monitor detected 61 arrhythmias compared to 52 arrhythmias detected by the patch monitor (P=.013). In review of the arrhythmia types three events were considered ‘clinically significant’ and were ultimately detected by the patch monitor beyond the first 24 hours.

In conclusion, the first seven days of AECG monitoring has the greatest detection of arrhythmias. Typically monitoring greater than seven days may only provide an additional 3.9% chance of identifying a different diagnosis as noted by the Zimetbaum (1998) article.

Rosenberg et al16 conducted a single-center pilot study to compare the use of a continuous LT-ECG recorder, with a 24-hour Holter monitor in 74 patients with paroxysmal A-Fib (AF), between April 2011 and May 2012. During the initial 24-hour monitoring period, both the Holter monitor, and the continuous LT-ECG recorder detected 25 AF occurrences. Patients continued to wear the continuous LT-ECG recorder for an additional two weeks, during which 43 more patients were diagnosed with AF. This indicates that AF episodes were detected in significantly more patients wearing a continuous LT-ECG recorder due to the longer duration of detection, compared with the Holter monitor. Furthermore, in this study, the LT-ECG recorder also detected cardiac pauses after the initial 24-hour monitoring, thus prompting physicians to review medications and initiate pacemaker referrals for this patient group.

Holter Monitor

Paudel et al17 conducted a prospective, single center study that included 335 patients who had symptoms of recurrent chronic palpitations or a single episode of palpitations. These patients were older than 18 years of age and the study was conducted from January 2010 to May 2012.

Patients were provided a Cardio Blue® 24-hour Holter monitor that recorded in either three or five channels and allowed for recording of arrhythmias, as defined by the American College of Cardiology (ACC)/American Heart Association (AHA), with or without corresponding documented symptoms.

The arrhythmias that were identified via the Holter monitor were: less than 10% ventricular ectopy, 36.7% bigeminy, less than 10% supraventricular episodes, 5.7% non-sustained VT and 12.5% sustained VT.

The study groups were noted to be statistically significant for patients that were younger than 50 years of age and those who were 50 years of age or older. The highest categories of arrhythmias were related to patients that were older than 50 years of age. This study group had a 72% rate with premature ventricular contactions (PVCs) and 44% with bigeminy as compared to 38.7% having PVCs and 23.1% with bigeminy that were younger than 50 years of age.

The authors reviewed literature provided by Mayet (1995) and Stein (2010). The conclusion of those two articles determined that Holter monitors detect arrhythmias such as ventricular bigeminy, VT, SVT, AV blocks, and silent ischemia in older patients who are more at risk of sudden cardiac death.

Event Recorders/Intermittent Monitors

Fredriksson et al18 studied intermittent AECG recordings versus continuous event recording in detecting A-Fib for patients 75-76 years of age. Two-hundred and sixty-nine patients that were included in this analysis were from the STROKESTOP II study from Europe. This study was conducted over a two-week period with 55 days on average for the intermittent recorders and 13 days on average for the continuous recorders. The devices were R-test 4® and Novocaor® (for continuous recording) and a 30 second handheld device Zenicor II® (for intermittent recording). All devices had activation buttons for symptomatic arrhythmias, and patients were given a symptom diary as well. Within the study, A-Fib was defined by the European Society of Cardiology (ESC) as “absolute irregular rate-to-rate intervals, no discernable distinct p-waves, and duration of at least 30 seconds” and recordings were manually inspected and validated on computerized algorithms.18(p360)

After the review of data, it was determined that continuous event recordings identified three times more cases of A-Fib (6%; n=15/269) than the intermittent AECG recordings (2%; n=5/269) (p=.002). On average, continuous recordings identified a 1-8 interquartile range (IQR) on day four and a 4-14 IQR on day eight for intermittent monitoring (P=.135). These IQR’s were analyzed using the Mann-Whitney U test and Chi-square tests in relation to proportions. While comparisons between the two monitoring methods were performed, the use of the McNemar’s test resulted in P< .05, which was considered significant when testing for dichotomous variables and paired sample t-test for continuous variables.18(p356)

Surprisingly, only six percent of the patients reported symptoms within their diaries when having verified A-Fib and none of the patients had reported palpitations. Limitations of the study include: patients that were part of a larger study (STROKESTOP II) and were thus highly motivated and possibly healthier than peers, use of one-lead AECGs that potentially complicate the analysis of the p-waves, possible bias by misclassification case, and conflicts of interest for those study team members receiving grants from companies such as Bayer, Pfizer, Sanofi, AstraZeneca, and Medtronic.

External Loop Recorders/Mobile Cardiovascular Outpatient Telemetry

Favilla et al19 provides a retrospective cohort study of 227 patients with cryptogenic stroke (179 patients) or transient ischemic stroke (48 patients). These patients underwent a 28-day mobile cardiac outpatient telemetry (MCOT), post cryptogenic or transient ischemic stroke. Of the 277 patients, 14% had A-Fib detected on MCOT, 58% of which was ≥ 30 seconds in duration. It was determined that age and prior cortical or cerebellar infarctions were independent predictors of A-Fib.

Rothman et al20 provided a prospective clinical trial of randomized patients that were evaluated with the CardioNet® system in 17 multicenter facilities. Two-hundred and sixty-six patients who had palpitations, presyncope, and/or syncope were included. The goal was to confirm or exclude the etiology of arrhythmias that caused patient symptoms. All patients had previously undergone 24 hours of monitoring with a Holter monitor, which failed to provide diagnostic information. These patients were placed in two groups for 30 days of monitoring with either the MCOT or with an external looping event monitor (ELM).

During monitoring, clinically significant arrhythmias were detected in 55 (41%) patients with the MCOT versus 19 (15%) patients with the ELM (P< 0.001). For patients who had prior syncope or presyncope, clinically significant arrhythmias were detected in 52% of patients with MCOT and in 16% of patients with loop recorders (P< 0.001). In most cases, the arrhythmias detected were A-Fib, atrial flutter, or ventricular tachycardia. There were two subgroups noted in the study. For one subgroup of patients presenting with “syncope or presyncope, a diagnosis was made in 89% of MCOT subjects versus 69% of LOOP subjects (P=0.008)."20(p243) A second subgroup analysis was performed at the centers that used auto-triggered loop monitoring rather than patient-activated monitoring. For this subgroup, a definitive diagnosis was obtained for 88% of MCOT patients and only 46% (P=0.002) of the ELM patients; however, this subgroup analysis involved only 50 of the 266 patients.

Insertable Cardiac Monitor/Subcutaneous Cardiac Rhythm Monitors

Nadkarni21 provides an overview of subcutaneous cardiac rhythm monitors (SCRMs) with the high diagnostic yield for “cryptogenic stroke, syncope, palpitation and AF [A-Fib] Monitoring."21(p588) It has long been noted that there is no established guideline for the use of SCRMs and the rate of false-positive transmissions has been the ‘Achilles heel’ of AECG. The author determines that the indications for use, techniques used in implantation, programming, and connectivity determine the clinical benefits for patients. Within this overview the author cites the ABACUS prospective randomized control trial by Kapa et al. This trial noted that SCRMs “led to more actionable events and higher rates of antiarrhythmic discontinuation when compared with other forms of rhythm monitoring in a post AFib ablation population."21(p588)

The versatility of the SCRM allows for adjustments in tachycardia and bradycardia thresholds, minimum episode durations, and the detection of atrial tachycardia and atrial fibrillation. Along with the versatility is the ability to transmit data in two ways: alert notifications and manual downloads by the patient via their home or cell phone application. With that noted and continual advancements, SCRMs have proven to be beneficial for patients.

SCRMs remain a valuable tool in the diagnosis of cardiac events.

Professional Societies and Associations

American College of Cardiology (ACC)/American Heart Association (AHA) Task Force on Clinical Practice Guidelines

Kusumoto8 provides a consensus statement on behalf of the ACC/AHA and the task force for clinical practices. The statement provides an overview and recommendations regarding appropriate use of AECG monitoring, particularly in the context of diagnosing bradyarrhythmia, and patients with bradycardia and cardiac conduction delays, patients can be managed with a 24-to-48-hour continuous AECG. For patients that present with symptoms that are greater than 30 days between symptoms, the use of a broad array of modalities, such as long-term monitors, implantable cardiac monitors, etc., would be appropriate if the initial non-invasive evaluation was not diagnostically conclusive. The guideline indicates that the “Choice of device is predicated on frequency of symptoms and the degree to which symptoms incapacitate the patient” and that regardless of the monitoring system, it is important for the system to notify health care providers in a timely fashion, especially in regard to identifying potentially dangerous arrhythmias.8(p399)

American Heart Association /American College of Cardiology/ Heart Rhythm Society Clinical Practice Guidelines

Al-khatib9 provides clinical guidelines for the use of AECG monitoring indicating that suspected arrhythmias, the frequency of arrhythmias, and the associated symptoms dictate monitor type. For example, event or ‘Looping’ monitors are more appropriate and have greater diagnostic yields for sporadic symptoms. It further explains that when ventricular arrhythmias are identified and potentially harmful, AECG may not be appropriate. Further information upon the use of AECG monitoring is indicated when assessing medical therapy response.

International Society for Holter and Noninvasive Electrocardiology (ISHNE), and the Heart Rhythm Society (HRS) expert consensus statement

Steinberg7 provides a consensus statement that outlined the limitations, clinical indications, pharmacological treatment of arrhythmias, and the use of external monitoring for pacemaker malfunctioning/placement. It is noted that for the selection of specific AECG monitors, one must consider the diagnostic power, the monitoring capability and the accuracy, local availability, symptom frequency, patient compliance and the condition of the patient. Within the article there are descriptions of advantages and limitations associated with each AECG monitor. The clinical indications for use of the various types of monitors can include the following: syncope, bradyarrhythmia, tachyarrhythmia, palpitations, chest pain and coronary ischemia, ischemic heart diseases and postinfarction, hypertrophic cardiomyopathy, arrhythmic right ventricular dysplasia/cardiomyopathy, Wolff-Parkinson-White syndrome, inherited primary arrhythmic diseases, Short or Long QT syndrome, Brugada syndrome, catecholaminergic polymorphic ventricular tachycardias, early repolarization syndromes, idiopathic ventricular fibrillation with nonischemic dilated cardiomyopathy, dialysis and chronic kidney disease associated arrhythmias. It is noted that the use of monitoring with AECG for neurological and muscular disease is controversial. The article further indicates an appropriate clinical assessment may include a continuous AECG, and if unsuccessful, additional monitoring should include an intermittent external loop recording. If patients remain undiagnosed after prolonged noninvasive monitoring, an internal loop recorder (ILR) may be necessary.

Analysis of Evidence (Rationale for Determination)

Coverage of items and services in the Medicare program is provided based on what is medically reasonable and necessary. This concept is operationalized by the application of evidentiary standards, clinical validity, and clinical utility. Over the years, AECG monitoring has advanced and evolved to include a spectrum of technology, from the long standing Holter monitors to the present-day Bluetooth enabled AECG monitors that allow for increased functionality and patient convenience. Arrhythmias often remain unpredictable in onset, vary in duration and the episodes can be incapacitating and life threatening for the patient. The evidence thresholds for coverage are a careful balance of benefits and harms in the consideration of net health outcomes.

The FDA provides controls and guidance in relation to AECG’s through the FDA 510(k) device approval. There are over 400 approved AECG monitoring devices aimed at improving the recognition, interpretation, and etiology of various arrhythmias. With these devices, health care professionals have many choices for obtaining the necessary diagnostic information.

In review of these clinical trials, randomized controlled trials (RCTs), meta-analyses, societal recommendations, and literature it is noted that there are different standards for AECG based on patient symptoms, antiarrhythmic regulation, post internal cardiac defibrillator insertion, cryptogenic stroke, cardio-embolism, transient ischemic stroke, silent myocardial ischemia, conduction disorders, and the evaluation of A-Fib/A-Flutter. It is generally expected that through a detailed exam that is based on patient signs and symptoms, a thorough cardiac exam, review of the patient medical/family histories to assess for conduction system abnormalities, valvular disease, metabolic disorders, endocrine issues, chronic obstructive pulmonary disease, structural heart disease, toxicology, and physiological disorders, the practitioner will be able to diagnose and treat the patient. If further diagnostic information is needed, a 12-lead EKG is appropriate and when combined with the patient’s capability and potential compliance, will aid the healthcare provider in determining if AECG monitoring is appropriate and which AECG monitoring device should be used.

Societal evidence provides the support for the use of AECG when evaluating for effective use of antiarrhythmic drugs, in the context of pre/post operative re-programing of defibrillators, to assess for known or suspected coronary artery disease, for evaluation and assessment of asymptomatic ventricular premature beats, and for evaluation and assessment of non-sustained ventricular tachycardia in patients with hypertrophic cardiomyopathy, arrhythmogenic right ventricular cardiomyopathy, long QT syndrome, dilated or restrictive cardiomyopathy, congenital heart disease, or Brugada syndrome.

The progression of non-invasive diagnostics to more invasive diagnostics is the standard of care taking into consideration benefits versus harm to patients while determining the etiology and treatment options for patients with arrhythmias. With the advancements in monitoring capabilities the ordering provider has multiple available alternatives to determine the type of monitoring needed for the individual needs of the patient.

After review of the available literature, it has been determined that limited coverage for AECG is medically reasonable and necessary when a 12-lead EKG does not provide an adequate diagnosis. It is expected that providers will use the most appropriate AECG device consistent with the patients’ signs and symptoms.

Proposed Process Information

Synopsis of Changes
Changes Fields Changed
N/A
Associated Information
Sources of Information
Bibliography
Open Meetings
Meeting Date Meeting States Meeting Information
N/A
Contractor Advisory Committee (CAC) Meetings
Meeting Date Meeting States Meeting Information
N/A
MAC Meeting Information URLs
N/A
Proposed LCD Posting Date
Comment Period Start Date
Comment Period End Date
Reason for Proposed LCD
Requestor Information
This request was MAC initiated.
Requestor Name Requestor Letter
View Letter
N/A
Contact for Comments on Proposed LCD

Coding Information

Bill Type Codes

Code Description

Please accept the License to see the codes.

N/A

Revenue Codes

Code Description

Please accept the License to see the codes.

N/A

CPT/HCPCS Codes

Please accept the License to see the codes.

N/A

ICD-10-CM Codes that Support Medical Necessity

Group 1

Group 1 Paragraph:

N/A

Group 1 Codes:

N/A

N/A

ICD-10-CM Codes that DO NOT Support Medical Necessity

Group 1

Group 1 Paragraph:

N/A

Group 1 Codes:

N/A

N/A

Additional ICD-10 Information

General Information

Associated Information

Please refer to the related Local Coverage Article: Billing and Coding: Ambulatory Electrocardiograph (AECG) Monitoring, A59270 for documentation requirements, utilization parameters and all coding information as applicable.

Sources of Information
N/A
Bibliography

This bibliography presents those sources that were obtained during the development of this policy. The Contractor is not responsible for the continuing viability of Website addresses listed below.

  1. Scirica BM. Acute coronary syndrome: emerging tools for diagnosis and risk assessment. J Am Coll Cardiol. 2010;55(14):1403-1415. doi:10.1016/j.jacc.2009.09.071.
  2. Cohn P, Fox KM, Daly C. Silent Myocardial Ischemia. Circulation. 2003;108:1263–1277. doi.org/10.1161/01.CIR.0000088001.
  3. National Coverage Determination (NCD) for Electrocardiographic Services (20.15) Centers for Medicare & Medicaid Services. 2022. https://www.cms.gov/medicare-coverage-database/view/ncd.aspx?NCDId=179. Accessed August 8, 2022.
  4. Decision Memo for Electrocardiographic Services (CAG-00158N) August 26, 2004. Centers for Medicare and Medicaid Services.
  5. Technology Assessment Report Project ID: CRDT0511. June 26, 2013. Agency for Healthcare Research and Quality.
  6. Madias C, Zimetbaum P, Parikh N. Ambulatory ECG Monitoring. 2020. https://www.wolterskluwer.com/en/solutions/uptodate. Accessed August 6, 2022.
  7. Steinberg JS, Varma N, Cygankiewicz I, et al. 2017 ISHNE-HRS expert consensus statement on ambulatory ECG and external cardiac monitoring/telemetry. Ann Noninvasive Electrocardiol. 2017;22:e12447. doi:10.1111/anec.12447.
  8. Kusumoto FM, Schoenfeld MH, Barrett C, et al. 2018 ACC/AHA/HRS Guideline on the Evaluation and Management of Patients with Bradycardia and Cardiac Conduction Delay. Circulation. 2019;140:e382–e482. doi:10.1161/CIR.0000000000000628.
  9. Al-Khatib SM, Stevenson WG, Ackerman MJ, et al. 2017 AHA/ACC/HRS Guideline for Management of Patients With Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death: A Report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society [published correction appears in J Am Coll Cardiol. 2018 Oct 2;72(14):1760]. J Am Coll Cardiol. 2018;72(14):e91-e220. doi:10.1016/j.jacc.2017.10.054.
  10. Patel UK, Malik P, Patel N, et al. Newer Diagnostic and Cost-Effective Ways to Identify Asymptomatic Atrial Fibrillation for the Prevention of Stroke. Cureus. 2021:13(1),e12437. doi:10.7759/cureus.12437.
  11. Varma N, Cygankiewicz I, Turakhia M, et al. 2021 ISHNE/HRS/EHRA/APHRS collaborative statement on mHealth in arrhythmia management: digital medical tools for heart rhythm professionals: from the International Society for Holter and Noninvasive Electrocardiology/Heart Rhythm Society/European Heart Rhythm Association/Asia Pacific Heart Rhythm Society. J Arrhythm. 2021;37(2):271-319.
  12. Sharma AN, Baranchuk A. Ambulatory External Electrocardiography Monitoring: Holter, Extended Holter, Mobile Cardiac Telemetry Monitoring. Card Electrophysiol Clin. 2021;13:427–438. doi:10.1016/j.ccep.2021.04.003.
  13. Zimetbaum P, Goldman A. Ambulatory Arrhyhmia Monitoring. Circulation. 2010;122:1629-1636. doi: 10.1161/CIRCULATIONAHA.109.925610.
  14. Zimetbaum P, Aronson M, Givens J. Evaluation of palpitations in adults. UpToDate. 2021;Sept. www.uptodate.com.
  15. Barrett PM, Komatireddy R, Hasser S, et al. Comparison of 24-hour Holter Monitoring with 14-day Novel Adhesive Patch Electrocardiographic Monitoring. The American Journal of Medicine. 2014;127:95.e11-95.e17.
  16. Rosenberg MA, Samuel M, Thosani A, Zimetbaum PG. Use of a Noninvasive Continuous Monitoring Device in the Management of Atrial Fibrillation: A Pilot Study. Wiley Periodicals, Inc. 2013;36. doi:10.1111/pace.12053.
  17. Paudel B, Paudel K. The diagnostic significance of the Holter monitoring in the evaluation of palpitation. JCDR. 2013;7(3):480–483. doi:10.7860/JCDR/2013/4923.2802.
  18. Fredriksson T, Gudmundsdottir KK, Frykman V, et al. Intermittent vs continuous electrocardiogram event recording for detection of atrial fibrillation—Compliance and ease of use in an ambulatory elderly population. Clinical Cardiology. 2020;43:355-362. doi:10.1002/clc.23323.
  19. Favilla CG, Ingala E, Jara J, et al. Predictors of finding occult atrial fibrillation after cryptogenic stroke. Stroke. 2015;46(5):1210-5.
  20. Rothman SA, Laughlin JC, Seltzer J, et al. The diagnosis of cardiac arrhythmias: A prospective multi-center randomized study comparing mobile cardiac outpatient telemetry versus standard loop event monitoring. J Cardiovasc Electrophysiol. 2007;18(3):241-247.
  21. Nadkarni A, Devgun J, Jamal SM, et al. Subcutaneous cardiac rhythm monitors: state of the art review. Expert Review of Medical Devices. 2021;18(7):587-596. doi:10.1080/17434440.2021.1935873.
  22. Bocchiardo M, Asteggiano R. ECG portable devices: example of e-Health strength and threats. ESC. 2020:(18):1-13.
  23. Hoppe B, Rees A, Parmar S, et al. Syncope pathway using live ambulatory monitoring streamlines ER patient disposition. J Am Coll Cardiol. 2022;79(9)A. doi:10.1016/S0735-1097(22)02989-8.
  24. Kishore A, Vail A, Majid A, et al. Detection of atrial fibrillation after ischemic stroke or transient ischemic attack: a systematic review and meta-analysis. Stroke. 2014;45(2):520-6.
  25. Kadish AH, Reiffel JA, Clauser J, et al. Frequency of serious arrhythmias detected with ambulatory cardiac telemetry. Am J Cardiol. 2010;105(9):1313-6.
  26. Locati ET, Vecchi AM, Vargiu S, et al. Role of extended external loop recorders for the diagnosis of unexplained syncope, pre-syncope, and sustained palpitations. Europace. 2014;16(6):914-922. doi:10.1093/europace/eut337.
  27. Lui CM, Chang SL, Yeh YH, et al. Enhanced detection of cardiac arrhythmias utilizing 14-day continuous ECG patch monitoring. International Journal of Cardiology. 2021;332:78-84. doi:10.1016/j.ijcard.2021.03.015.
  28. Sana F, Isselbacher EM, Singh JP, et al. Wearable Devices for Ambulatory Cardiac Monitoring JACC State-of-the-Art Review. Journal of the American College of Cardiology. 2020;75(13)1582-1592. doi:10/1016/j.jacc.2020.01.0476.
  29. Sciaraffia E, Chen J, Hocini M, et al. Use of event recorders and loop recorders in clinical practice: results of the European Heart Rhythm Association Survey. EP Europace. 2014;16(9):1384–1386. doi:10.1093/europace/euu222.
  30. Singer DE, Atlas S, Go AS, Lopes RD, et al. A Randomized clinical trial of screening for atrial fibrillation with a 14-day patch monitor: Analysis of ECG recording from the Guard-AF Study. JACC. 2022;79(9):28.
  31. Tan ESJ, Seow SC, Pipin K, et al. Optimal duration and predictors of diagnostic utility of patient-activated ambulatory ECG monitoring. Heart Asia. 2018;10:e011061. doi:10.1136/heartasia-2018-011061.
  32. January CT, Wann LS, Calkins H, et al. 2019 AHA/ACC/HRS Focused Update of the 2014 AHA/ACC/HRS Guideline for the Management of Patients With Atrial Fibrillation: A Report of the American College of Cardiology/American Heart AssociationTask Force on Clinical Practice Guidelines and the Heart Rhythm Society. Circulation. 2019;140:e125-e151. doi:10.1161/CIR.0000000000000665.

Revision History Information

Revision History Date Revision History Number Revision History Explanation Reasons for Change
N/A

Associated Documents

Attachments
N/A
Related National Coverage Documents
N/A
Public Versions
Updated On Effective Dates Status
04/21/2023 06/11/2023 - N/A Currently in Effect You are here

Keywords

N/A

Read the LCD Disclaimer