National Coverage Determination (NCD)

Implantable Automatic Defibrillators

20.4

Expand All | Collapse All

Tracking Information

Publication Number
100-3
Manual Section Number
20.4
Manual Section Title
Implantable Automatic Defibrillators
Version Number
3
Effective Date of this Version
01/27/2005
Ending Effective Date of this Version
02/15/2018
Implementation Date
01/27/2005
Implementation QR Modifier Date

Description Information

Benefit Category
Prosthetic Devices


Please Note: This may not be an exhaustive list of all applicable Medicare benefit categories for this item or service.

Item/Service Description

A. General

The implantable automatic defibrillator is an electronic device designed to detect and treat life-threatening tachyarrhythmias. The device consists of a pulse generator and electrodes for sensing and defibrillating.

Indications and Limitations of Coverage

B. Covered Indications

  1. Documented episode of cardiac arrest due to ventricular fibrillation (VF), not due to a transient or reversible cause (effective July 1, 1991).
  2. Documented sustained ventricular tachyarrhythmia (VT), either spontaneous or induced by an electrophysiology (EP) study, not associated with an acute myocardial infarction (MI) and not due to a transient or reversible cause (effective July 1, 1999).
  3. Documented familial or inherited conditions with a high risk of life-threatening VT, such as long QT syndrome or hypertrophic cardiomyopathy (effective July 1, 1999).
  4. Additional indications effective for services performed on or after October 1, 2003:

  5. Coronary artery disease with a documented prior MI, a measured left ventricular ejection fraction (LVEF) ≤ 0.35, and inducible, sustained VT or VF at EP study. (The MI must have occurred more than 40 days prior to defibrillator insertion. The EP test must be performed more than 4 weeks after the qualifying MI.)
  6. Documented prior MI and a measured LVEF ≤ 0.30 and a QRS duration of >120 milliseconds  (the QRS restriction does not apply to services performed on or after January 27, 2005) . Patients must not have:
    1. New York Heart Association (NYHC) classification IV;
    2. Cardiogenic shock or symptomatic hypotension while in a stable baseline rhythm;
    3. Had a coronary artery bypass graft (CABG) or percutaneous transluminal coronary angioplasty (PTCA) within past 3 months;
    4. Had an enzyme positive MI within past month  (Effective for services on or after January 27, 2005, patients must not have an acute MI in the past 40 days);
    5. Clinical symptoms or findings that would make them a candidate for coronary revascularization; or
    6. Any disease, other than cardiac disease (e.g., cancer, uremia, liver failure), associated with a likelihood of survival less than 1 year.

    Additional indications effective for services performed on or after January 27, 2005:

  7. Patients with ischemic dilated cardiomyopathy (IDCM), documented prior MI, NYHA Class II and III heart failure, and measured LVEF ≤ 35%;
  8. Patients with non-ischemic dilated cardiomyopathy (NIDCM) >9 months, NYHA Class II and III heart failure, and measured LVEF ≤ 35%;
  9. Patients who meet all current Centers for Medicare & Medicaid Services (CMS) coverage requirements for a cardiac resynchronization therapy (CRT) device and have NYHA Class IV heart failure;

    All indications must meet the following criteria:

    1. Patients must not have irreversible brain damage from preexisting cerebral disease;
    2. MIs must be documented and defined according to the consensus document of the Joint European Society of Cardiology/American College of Cardiology Committee for the Redefinition of Myocardial Infarction; 1

    Indications 3 - 8 (primary prevention of sudden cardiac death) must also meet the following criteria:

    1. Patients must be able to give informed consent;
    2. Patients must not have:
      • Cardiogenic shock or symptomatic hypotension while in a stable baseline rhythm;
      • Had a CABG or PTCA within the past 3 months;
      • Had an acute MI within the past 40 days;
      • Clinical symptoms or findings that would make them a candidate for coronary revascularization;
      • Any disease, other than cardiac disease (e.g., cancer, uremia, liver failure), associated with a likelihood of survival less than 1 year;

    3. Ejection fractions must be measured by angiography, radionuclide scanning, or echocardiography;
    4. The beneficiary receiving the defibrillator implantation for primary prevention is enrolled in either a Food and Drug Administration (FDA)-approved category B investigational device exemption (IDE) clinical trial (42 CFR §405.201), a trial under the CMS Clinical Trial Policy (National Coverage Determination (NCD) Manual §310.1) or a qualifying data collection system including approved clinical trials and registries. Initially, an implantable cardiac defibrillator (ICD) database will be maintained using a data submission mechanism that is already in use by Medicare participating hospitals to submit data to the Iowa Foundation for Medical Care (IFMC) a Quality Improvement Organization (QIO) contractor for determination of reasonable and necessary and quality improvement. Initial hypothesis and data elements are specified in this decision (Appendix VI) and are the minimum necessary to ensure that the device is reasonable and necessary. Data collection will be completed using the ICDA (ICD Abstraction Tool) and transmitted via QNet (Quality Network Exchange) to the IFMC who will collect and maintain the database. Additional stakeholder-developed data collection systems to augment or replace the initial QNet system, addressing at a minimum the hypotheses specified in this decision, must meet the following basic criteria:

      • Written protocol on file;
      • Institutional review board review and approval;
      • Scientific review and approval by two or more qualified individuals who are not part of the research team;
      • Certification that investigators have not been disqualified.

      For purposes of this coverage decision, CMS will determine whether specific registries or clinical trials meet these criteria.

    5. Providers must be able to justify the medical necessity of devices other than single lead devices. This justification should be available in the patient's medical record.

  10. Patients with NIDCM >3 months, NYHA Class II or III heart failure, and measured LVEF ≤ 35%, only if the following additional criteria are also met:

    1. Patients must be able to give informed consent;
    2. Patients must not have:

      • Cardiogenic shock or symptomatic hypotension while in a stable baseline rhythm;
      • Had a CABG or PTCA within the past 3 months;
      • Had an acute MI within the past 40 days;
      • Clinical symptoms or findings that would make them a candidate for coronary revascularization;
      • Irreversible brain damage from preexisting cerebral disease;
      • Any disease, other than cardiac disease (e.g. cancer, uremia, liver failure), associated with a likelihood of survival less than 1 year;

    3. Ejection fractions must be measured by angiography, radionuclide scanning, or echocardiography;
    4. MIs must be documented and defined according to the consensus document of the Joint European Society of Cardiology/American College of Cardiology Committee for the Redefinition of Myocardial Infarction; 2
    5. The beneficiary receiving the defibrillator implantation for this indication is enrolled in either an FDA-approved category B IDE clinical trial (42 CFR §405.201), a trial under the CMS Clinical Trial Policy (NCD Manual §310.1), or a prospective data collection system meeting the following basic criteria:

      • Written protocol on file;
      • Institutional Review Board review and approval;
      • Scientific review and approval by two or more qualified individuals who are not part of the research team;
      • Certification that investigators have not been disqualified.

      For purposes of this coverage decision, CMS will determine whether specific registries or clinical trials meet these criteria.

    6. Providers must be able to justify the medical necessity of devices other than single lead devices. This justification should be available in the patient's medical record.

C. Other Indications

All other indications for implantable automatic defibrillators not currently covered in accordance with this decision will continue to be covered under Category B IDE trials (42 CFR §405.201) and the CMS routine clinical trials policy (NCD §310.1).

(This NCD last reviewed February 2005.)


1 Alpert and Thygesen et al., 2000. Criteria for acute, evolving or recent MI.

Either one of the following criteria satisfies the diagnosis for an acute, evolving or recent MI:

  1. Typical rise and gradual fall (troponin) or more rapid rise and fall (CK-MB) of biochemical markers of myocardial necrosis with at least one of the following:

    1. ischemic symptoms;
    2. development of pathologic Q waves on the ECG;
    3. ECG changes indicative of ischemia (ST segment elevation or depression); or
    4. coronary artery intervention (e.g., coronary angioplasty).

  2. Pathologic findings of an acute MI.

Criteria for established MI.

Any one of the following criteria satisfies the diagnosis for established MI:

  1. Development of new pathologic Q waves on serial ECGs. The patient may or may not remember previous symptoms. Biochemical markers of myocardial necrosis may have normalized, depending on the length of time that has passed since the infarct developed.
  2. Pathologic findings of a healed or healing MI.

2 Ibid

Cross Reference

Transmittal Information

Transmittal Number
29
Revision History

06/2016 - Transmittal 1665, dated May 13, 2016, is being rescinded and replaced by Transmittal 1672 to: (1) 9631.1: Remove FISS responsibility and 1st sentence; (2) 9631.2: Remove additional procedure codes, including 0075T, 0076T; (3) 9631.4: Revise descriptor of dx L59.8; (4) 9631.6: Add deletion of dx C49.10, C65.9, remove deletion of dx C54.9, remove deletion of invalid dx C47.90; (5) 9631.8: Remove deletion of invalid dx C51.29, replace with deletion of dx C50.029; (6) 9631.9: Add deletion of 0V504ZZ, 0V500ZZ, override capability, and contractor discretion verbiage. All other information remains the same. (TN 1672) (CR9631)

05/2016 - This change request (CR) is the 7th maintenance update of ICD-10 conversions and other coding updates specific to national coverage determinations (NCDs). The majority of the NCDs included are a result of feedback received from previous ICD-10 NCD CR7818, CR8109, CR8197, CR8691, CR9087, CR9252, and CR9540. Some are the result of revisions required to other NCD-related CRs released separately.
Edits to ICD-10 and other coding updates specific to NCDs will be included in subsequent, quarterly releases as needed. No policy-related changes are included with these updates. Any policy-related changes to NCDs continue to be implemented via the current, long-standing NCD process. (TN 1665) (CR9631)

03/2013 - CMS translated the information for this policy from ICD-9-CM/PCS to ICD-10-CM/PCS according to HIPAA standard medical data code set requirements and updated any necessary and related coding infrastructure. These updates do not expand, restrict, or alter existing coverage policy. Implementation date: 10/07/2013 Effective date: 10/1/2015. (TN 1199) (TN 1199) (CR 8197)

3/2005 - Expanded national coverage by including new indications. Effective and implementation dates 01/27/2005. (TN 29) (CR 3604)

08/2003 - Expanded coverage for coronary artery disease with a documented prior myocardial infarction (MI), a measured left ventricular ejection fraction ≤ 0.35, and inducible, sustained ventricular tachyarrhythmia or ventricular fibrillation at electrophysiology study; or documented prior MI and a measured left ventricular ejection fraction ≤0.30 and a QRS duration of >120 milliseconds. All other indications for defibrillators not otherwise specified remain noncovered except when furnished in accordance with FDA-approved protocols governing Category B Investigational Device Exemption; clinical trials as stated in 42 FR Part 405. Effective and implementation dates 10/01/2003. (TN 173) (CR 2880)

04/1999 - Expanded coverage for additional types of cardiac patients. Effective and implementation dates 07/01/1999. (TN 115) (CR 825)

01/1986 - Provided coverage to treat life-threatening ventricular tachyarrhythmias. Effective date 01/24/1986. (TN 3)

Other

Coding Analyses for Labs (CALs)

This NCD has been or is currently being reviewed under the National Coverage Determination process. The following are existing associations with CALs, from the Coding Analyses for Labs database.

Additional Information

Other Versions
Title Version Effective Between
Implantable Cardioverter Defibrillators (ICDs) 5 07/31/2023 - N/A View
Implantable Cardioverter Defibrillators (ICDs) 4 02/15/2018 - 07/31/2023 View
Implantable Automatic Defibrillators 3 01/27/2005 - 02/15/2018 You are here
Implantable Automatic Defibrillators 2 10/01/2003 - 01/27/2005 View
Implantation of Automatic Defibrillators 1 07/01/1999 - 10/01/2003 View
CPT Copyright Statement
CPT only copyright 2002-2011 American Medical Association. All rights reserved. CPT is a registered trademark of the American Medical Association. Applicable FARS/DFARS Apply to Government Use. Fee schedules, relative value units, conversion factors and/or related components are not assigned by the AMA, are not part of CPT, and the AMA is not recommending their use. The AMA does not directly or indirectly practice medicine or dispense medical services. The AMA assumes no liability for data contained or not contained herein.
Reasons for Denial
Note: This section has not been negotiated by the Negotiated RuleMaking Committee. It includes CMS’s interpretation of it’s longstanding policies and is included for informational purposes. Tests for screening purposes that are performed in the absense of signs, symptoms, complaints, or personal history of disease or injury are not covered except as explicity authorized by statue. These include exams required by insurance companies, business establishments, government agencies, or other third parties. Tests that are not reasonable and necessary for the diagnosis or treatment of an illness or injury are not covered according to the statue. Failure to provide documentation of the medical necessity of tests may result in denial of claims. The documentation may include notes documenting relevant signs, symptoms, or abnormal findings that substantiate the medical necessity for ordering the tests. In addition, failure to provide independent verification that the test was ordered by the treating physician (or qualified nonphysician practitioner) through documentation in the physician’s office may result in denial. A claim for a test for which there is a national coverage or local medical review policy will be denied as not reasonable and necessary if it is submitted without an ICD-9-CM code or narrative diagnosis listed as covered in the policy unless other medical documentation justifying the necessity is submitted with the claim. If a national or local policy identifies a frequency expectation, a claim for a test that exceeds that expectation may be denied as not reasonable and necessary, unless it is submitted with documentation justifying increased frequency. Tests that are not ordered by a treating physician or other qualified treating nonphysician practitioner acting within the scope of their license and in compliance with Medicare requirements will be denied as not reasonable and necessary. Failure of the laboratory performing the test to have the appropriate Clinical Laboratory Improvement Act of 1988 (CLIA) certificate for the testing performed will result in denial of claims.