Policy Clarification on Acceptable Control Materials Used when Quality Control (QC) is Performed in Laboratories

Dynamic List Information
Dynamic List Data
Title
Policy Clarification on Acceptable Control Materials Used when Quality Control (QC) is Performed in Laboratories
Memo #
16-20-CLIA
Posting Date
2016-04-08
Fiscal Year
2016
Title
The Centers for Medicare & Medicaid Services (CMS) is providing clarification regarding the policy on acceptable control materials, which includes:
• Acceptable control materials: Acceptable control materials will now include on-board controls, i.e. ampules or cartridges containing the same QC material that would traditionally be considered as external QC.
• Function checks and procedural controls: Instrument/electronic function checks and procedural controls do not fulfill the regulatory requirement for control materials.
• Guidance for surveyors: The laboratory Director is responsible for the determination of what control materials to use in his/her laboratory. Surveyors will ensure that the laboratory is following its own established policies, specifically its QC procedures, in the context of the Outcome Oriented Survey Process.